1. Z6·尊龙凯时「中国」官方网站

    Product catalog
    • 1/1
    Previous
    Contact Us
    Online Message
    Detail

    the neonatal/pediatric continuous positive airway pressure breathing support system is a positive pressure breathing support system that uses a nasal mask or nasal congestion to send a continuous positive pressure airflow into the airway. Its principle is to use medical air and oxygen sources as power sources to output mixed Oxygen concentration, flow adjustable, controlled maximum pressure (MAX-P), peak inspiratory pressure (PIP), CPAP/PEEP pressure equipment, suitable for neonates and children with spontaneous breathing, providing continuous positive pressure ventilation support. The device is a manually operated, gas-driven breathing support system.

    Product advantages

    1. Controlled peak inspiratory pressure ( PIP ) to safely expand the lungs of newborns/children and provide oxygenation;   

    2. Constant CPAP/PEEP to keep the lungs open during exhalation, thereby avoiding The adhesion is damaged and the functional residual capacity is maintained;   

    3. It is easy to operate with one hand, and the operator is not fatigued by operating the bladder;   

    4. The respiratory support system can replace temporary breathing and resuscitation support;   

    5. The operator's experience, training, and attention Neither strength nor fatigue will affect the output pressure;   

    6. Constant CPAP/PEEP can help babies breathe while being transported or when changing ventilator tubes;

    7. It is mainly used as a resuscitation device for newborns and children with a weight of up to 10 kg in delivery rooms, infant wards and neonatal intensive care units to provide respiratory support.

    Technical specifications


    model

    AD3000-TPB

    » Flow Regulation LPM

    0 LPM-15LPM

    » Oxygen concentration O2%

    21%-100%

    » Air supply failure and failure alarm

    Listen to sense alarm

    » Pressure Gauge

    -2-10kPa ( -20~100cmH2O)

    » Pressure Limit ( MAX-P )

    @ 0-15 LPM ≤ 6kPa(60 cmH2O )

    » Peak Inspiratory Pressure ( PIP )

    @ At 5-15 the LPM of 0. The 2 - 5. The 7kPa (2-57 cmH2O )

    » CPAP/ PEEP

    @ 5-15 LPM 0.03 - 2.3 kPa (0.3-23 cmH2O )

    » Work noise

    Not more than 55dB ( A )

    » Air source requirements

    0.3MPa-0.4MPa capacity greater than 100L

    » Safety / Performance Requirements

    YY0600.5-2001/YY0893-2013

    » Recommended applicable weight range

    Heaviest 10KG








    Website links
    >1688 alibaba
    > alibaba.com

    > 3618med

    > merchants recruiting net

    >Made in China

    >MedicalExpo en
    >MedicalExpo cn